Intended Use vs. Intended Purpose

Regulatory Affairs Bulletin: Intended Purpose vs Intended Use under MDR & IVDR

European regulations EU 2017/745 (MDR) and EU 2017/746 (IVDR) require precise definitions in technical documentation, and the distinction (or overlap) between intended purpose and intended use remains a topic of operational significance. This bulletin clarifies the regulatory definitions, their implications, and guidance interpretations.

Definitions under MDR & IVDR

MDR (EU 2017/745), Article 2(12):

“‘intended purpose’ means the use for which a device is intended according to the data supplied by the manufacturer on the label, in the instructions for use or in promotional or sales materials or statements and as specified by the manufacturer in the clinical evaluation.” (MDR)

IVDR (EU 2017/746), Article 2(12):

“‘intended purpose’ means the use for which a device is intended according to the data supplied by the manufacturer on the label, in the instructions for use or in promotional or sales materials or statements or as specified by the manufacturer in the performance evaluation.” (IVDR)

Key Takeaway: Intended purpose is a legal term. It is the manufacturer’s declared scope of what the device is meant to do, forming the foundation for regulatory classification, clinical evaluation, and conformity assessment.

Intended Use:

The term intended use is not explicitly defined in either MDR or IVDR. Regulatory and standards-based documents commonly treat intended use as a synonym of intended purpose. For example, the Medical Device Coordination Group guidance MDCG 2020-6 states:

“The MDR defines ‘intended purpose’, but not ‘intended use’. ‘Intended use’ should be considered to have the same meaning as ‘intended purpose’.” (MDCG 2020-09) Also, international standards such as ISO 14971:2019 use “intended use/intended purpose” interchangeably, with notes identifying typical elements (patient population, mode of action, environment, etc.).

In practice, intended use is often interpreted as the clinical or practical application of the device: how, where, and by whom the device is meant to be used. For example:

§  A diagnostic assay’s intended purpose may be “qualitative detection of virus X in human serum,” while its intended use specifies whether it is for professional laboratory use or point-of-care testing.

§  A surgical instrument’s intended purpose may state it is “for cutting tissue,” while intended use clarifies the clinical context, such as “for laparoscopic abdominal surgery.”

Key Takeaway: Intended use reflects the practical deployment of the device, often with more granularity than the legally binding intended purpose or it is used as a synonym of intended purpose.

Key Regulatory Implications

  • Classification:
    Under IVDR Annex VIII (classification rules), the intended purpose specified by the manufacturer is the controlling basis for assigning a device’s risk class. Ambiguities or broad statements can lead to higher class assignments.
  • Clinical / Performance Evaluation:
    MDR requires that clinical evaluation verify safety and performance, “when used as intended by the manufacturer,” which ties back to the declared intended purpose. IVDR similarly requires performance evaluation consistent with intended purpose.
  • General Safety and Performance Requirements (GSPRs):
    The GSPRs in Annex I of both MDR and IVDR refer to “intended purpose” in respect of safety, design, usability, labeling, etc. All downstream documentation (e.g. IFU, promotional materials) must align with that purpose. Inconsistency may result in non-compliance.
  • Avoiding Misleading Claims & Liability:
    Article 7 MDR / IVDR prohibits misleading information; if promotional materials or claims exceed or differ from the declared intended purpose, this may violate regulatory obligations.

Best Practices for Regulatory Affairs Professionals

From the definitions and guidance, the following practical recommendations emerge:

Definition & Documentation

Define the intended purpose early in product development; Use the wording of MDR/IVDR Article 2(12) as baseline.

Consistency

Ensure intended purpose is consistent across all documents: label, IFU, promotional material, clinical or performance evaluation, technical file (Annex II). Any divergence may be treated as an unapproved claim or misrepresentation.

Clarity vs Flexibility

Be specific enough to allow correct classification, risk assessment and evidence generation, while allowing legitimate future expansion (if desired) without excessively restrictive wording.

Use of “Intended Use”

Although not defined, when using “intended use” internally or in standards, clarify that it is equivalent to intended purpose. Avoid using “intended use” in ways that imply extended or additional functions beyond what is declared in intended purpose.

Claims & Marketing

Review promotional materials carefully to ensure they do not introduce new intended purposes or uses not supported by clinical/performance evidence. If use cases outside the stated purpose are foreseeable, include limitations in IFU and technical documentation.

Summary

  • Intended purpose is a legal term explicitly defined in both MDR (Art. 2(12)) and IVDR (Art. 2(12)); intended use is not defined but is commonly treated as synonymous with intended purpose.
  • The declared intended purpose drives critical regulatory outcomes: classification, conformity assessment, clinical/performance evaluation, labelling, and marketing claims.
  • Regulatory strategy must ensure that all documentation aligns with the same intended purpose, and that any “use” beyond that purpose is either excluded or appropriately supported and declared.

Latest news

MHRA Introduces New Device Registration Fee: What Manufacturers Need to Know

MHRA Introduces New Device Registration Fee: What Manufacturers Need to Know

The UK’s Medicines and Healthcare products Regulatory Agency (MHRA) has updated its “DORS Fees Guidance 2.1”, introducing a new annual fee model for device registration.

EUDAMED’s first four modules become mandatory on 28 May 2026 - what this means for medical device manufacturers

EUDAMED’s first four modules become mandatory on 28 May 2026 - what this means for medical device manufacturers

The European Commission (EC) has formally declared that the first four modules of EUDAMED - the European Database on Medical Devices - will become mandatory as of 28 May 2026.

MedNet EC-REP GmbH to Exhibit at WHX 2026 in Dubai

MedNet EC-REP GmbH to Exhibit at WHX 2026 in Dubai

MedNet EC-REP GmbH is excited to announce its participation in WHX 2026, taking place from February 9–12 in Dubai. As a trusted European Authorized Representative and compliance partner for global medical device manufacturers, we look forward to presenting our comprehensive regulatory services to an international audience.

Implications of the Update to the Borderline & Classification Manual under MDR and IVDR

Implications of the Update to the Borderline & Classification Manual under MDR and IVDR

On 12 September 2025, the European Commission published Version 4 of the Manual on Borderline and Classification under Regulations (EU) 2017/745 and 2017/746, developed by the Borderline and Classification Working Group (BCWG).

MedNet EC-REP GmbH at MEDICA 2025 – Visit Us in Hall 3, Booth D43

MedNet EC-REP GmbH at MEDICA 2025 – Visit Us in Hall 3, Booth D43

MedNet EC-REP GmbH is pleased to announce that we will once again be exhibiting at MEDICA 2025 in Düsseldorf, the world’s leading trade fair for the medical technology industry.

EU eIFU Regulation Update Draft: What Medical Device Professionals Need to Know

EU eIFU Regulation Update Draft: What Medical Device Professionals Need to Know

The European Commission has introduced a significant amendment to Implementing Regulation (EU) 2021/2226, impacting how medical device manufacturers provide instructions for use (IFU).

Understanding the Interplay: MDR, IVDR, and the AI Act – Key Insights from MDCG 2025-6

Understanding the Interplay: MDR, IVDR, and the AI Act – Key Insights from MDCG 2025-6

Recently, the MDCG, in conjunction with the Joint Artificial Intelligence Board (AIB), released the MDCG 2025-6 FAQ document, offering essential clarifications on the interplay between the MDR, the IVDR, and the Artificial Intelligence Act (AIA).